The examination of minute objects by means of a
microscope, an instrument which provides an enlarged image of an object not
visible with the naked eye.
There are two fundamentally different types of
microscope: the light microscope and
the electron microscope.
ü Light
microscope (optical
microscope) involves the use of optical lenses and light radiations.
Light or optical microscopy involves passing visible
light transmitted through or reflected from the specimen through a single lens
or multiple lenses to allow a magnified view of the specimen.
The light
microscope creates a magnified image of the specimen which is based on the
principle of transmission, adsorption,
diffraction, and refraction of
light waves.
A simple microscope consists of a single lens, a magnifying glass.
The compound microscope consists of more than one glass lens in combination
and various components which gather light and redirects the light path so that
a magnified image of the viewed object can be focused within a short distance.
ü Functions
of the major parts of a light microscope-
Light source and condenser-
project a parallel beam of light onto the specimen for illumination.
Condenser gathers light and concentrates it into a
cone of light that illuminates the specimen with uniform intensity over the
entire view field.
Sample stage with X-Y
movement-sample is
placed on the stage and different part of the sample (specimen) can be viewed due
to the X-Y movement capability.
Fine/focusing knobs-since the distance between objective and eyepiece is
fixed, focusing is achieved by moving the sample relative to the objective
lens.
Objectives- does the main part of magnification and resolves the
fine details on the samples (mo-4X, 10X, 20X, 40X, 60X, 100X)
Eyepiece- forms a further magnified virtual image which can be
observed directly with eyes (me̴10)
ü Magnification
The magnification or linear magnification of a
microscope is defined as the ratio of the image size to the specimen size. To
visualize any tiny object 0.1mm there is a limit of the unaided human eye. To see microorganisms much
smaller than 0.1mm a system required convex lenses. When parallel rays pass
through convex lens get converged at a point called focal length (f) of the lens.
Hence, a lens with a shorter focal length will have higher magnification power.
In compound microscope it will be i.e. 10X, f=16mm; 40X, f=4mm; 100X, f=1.8mm.
The overall magnification is given as the product of
the lenses and the distance over which the image is projected:
M= (D*M1*M2)/250mm
Where D=projection (tube) length (usually 250mm), M1
and M2 magnification of objective and ocular lens respectively.
250mm=minimum distance of distinct vision for 20/20
eyes.
ü Resolution
The ability to distinguish between two closely spaced
points in a specimen. The limit of resolution is the closest distance between
two points at which the points still can be distinguished as separate entities.
Standard light microscopes have a lateral resolution limit of about 0.5µm for
routine analysis. In contrast, electron microscopes have a lateral resolution
of up to 1 nanometre (nm).
Magnification should be coupled with good resolution
to visualize small microorganisms, else magnification alone will produce an
inconclusive or blurred image.
Resolution is described mathematically by an equation
developed in the 1870s by Ernst Abbe,
a German physicist responsible for much of the optical theory underlying
microscope design.
The Abbe equation states that the minimal distance (d) between two objects that reveals them as separate
entities depend on the wavelength
of light (λ) used to illuminate the specimen and on the numerical aperture of the lens (nsinϴ),
which is the ability of the lens to gather light.
d= 0.5λ / nsinϴ
As d becomes smaller, the resolution increases, and
finer detail can be discerned in a specimen, d becomes smaller as the wavelength of light used decreases and as the numerical aperture increases.
Thus the greatest resolution is obtained using a lens with the largest possible
numerical aperture and light of the shortest wavelength, the light at the blue end
of the visible spectrum (in the range of 450nm to 500nm). The numerical
aperture (nsinϴ) of a lens is defined by two components: n is the refractive index of the medium in which the lens works
(e.g., air) and ϴ is ½ the angle of the
cone of light entering an objective. A cone with a narrow-angle does not
adequately separate the rays of light emanating from closely packed objects,
and the images are not resolved.
Numerical Aperture in
Microscopy- The angular
aperture ϴ is ½ the angle of the cone of light that enters a lens from a
specimen and the numerical aperture is nsinϴ. In the right-hand illustration,
the lens has larger angular and numerical apertures, its resolution is greater
and its working distance smaller.
A cone of light with a very wide angle does separate
the rays and the closely packed objects appear widely separated and resolved.
Some objective lenses work in air, which has a refractive index of 1.00. Since
sinϴ cannot be greater than 1 (the maximum ϴ is 90ͦ and sin 90ͦ is 1.00), no
lens working in the air can have a numerical aperture greater than 1.00.
The only practical way to raise the numerical aperture
above 1.00, and therefore achieve higher resolution, is to increase the
refractive index with immersion oil, a colorless liquid with the same
refractive index as glass.
ü Properties
of objective lenses
|
|
|
Objective |
|
|
|
Property |
Scanning |
Low Power |
High Power |
Oil immersion |
|
Magnification |
4X |
10X |
40X-45X |
90X-100X |
|
Numerical aperture |
0.10 |
0.25 |
0.55-0.65 |
1.25-1.4 |
|
Approximate focal length
(f) |
40mm |
16mm |
4mm |
1.8mm-2.0mm |
|
Working distance |
17mm-20mm |
4mm-8mm |
0.5mm-0.7mm |
0.1mm |
|
Approximate resolving
power with the light of 450 nm (blue light) |
2.3μm |
0.9μm |
0.35μm |
0.18μm |
If air is replaced with immersion oil, many light rays
that did not enter the objective due to reflection and refraction at the
surfaces of the objective lens and slide will now do so. This results in an
increase in numerical aperture and resolution.
The oil immersion objective- An oil immersion objective operating in air and with
immersion oil.
The limit of resolution of a light microscope is
calculated using the Abbe equation. The maximum theoretical resolving power of
a microscope when viewing a specimen using an oil immersion objective
(numerical aperture of 1.25) and blue-green light is approximately 0.2μm.
d= (0.5) (530nm) / 1.25
= 212nm or 0.2μm
At best, a light microscope can distinguish between
two dots about 0.2μm apart (the same size as a very small bacterium). Thus the
vast majority of viruses cannot be examined with a light microscope.
ü Bright-Field
Microscope: Dark object,
Bright background
The most basic model of the light microscope is called
bright-field (bright background), which can be achieved with a minimum of
optical elements. It is used to examine both stained and unstained specimens.
It is called a bright field microscope because it forms a dark image against a
brighter background.
The contrast in bright-field images is usually produced by
the color or dense sections of the specimen itself. Bright-field is therefore
used most often to collect images from pigmented sections of the specimen.
However, many microbes are unpigmented and are not clearly visible because
there is little difference in contrast between the cells, subcellular
structures, and water. One solution to this problem is to stain cells before
observation to increase contrast and create variations in color between cell
structures. Unfortunately, staining procedures usually kill cells.
Working
·
Bright
field microscopy is a technique used in the light microscope which gives a
magnified image of the dark specimen with a colorless background. To
accomplish the bright field microscopy, transfer the glass slide having a
stained specimen onto the microscope stage.
·
The
luminous light will come through the source of light or we can say through a
source of illumination. The light coming from the illuminator is aimed at the
condenser lens, which is present beneath the specimen. The aperture diaphragm
helps to focus and control the light coming from the light source illuminator on the specimen.
·
Some
of the light will reflect out of the specimen, which is then collected by the
objective lens. The objective lens first magnifies the light and then transmits
it to the ocular lens. During this whole method, some of the light will get
deflected, and some will get absorbed by the stain and the dense areas in the
specimen.
·
The
intense illumination and can increase the magnification of the image.
Therefore, in bright field microscopy, those which have absorbed the part of
the light will appear darker and the remaining, i.e. the background will appear
brighter.
Advantages
·
Bright
field microscopy is a simple method to perform, simplicity of setup with only
basic equipment required.
·
It
can easily produce a magnified image of the fixed specimens and live cells.
Disadvantages
·
The
bright field microscopy produces low contrast to the image.
·
The practical limit to magnification with a light microscope is around 1300X.
Although higher magnifications are possible, it becomes increasingly difficult
to maintain image clarity as to the magnification increases.
·
Low
apparent optical resolution due to the blur of out-of-focus material.
Enhancements
·
Reducing
and increasing the amount of the light source coming to the specimen by
adjusting the iris diaphragm.
·
Use
of an oil-immersion objective lens and a special immersion oil placed on a
glass cover over the specimen. Immersion oil has the same refraction as glass
and improves the resolution of the observed specimen.
·
The
use of staining methods adds contrast to the picture.
·
Use
of a colored (usually blue) or polarizing filter on the light source to
highlight features not visible under white light. The use of filters is
especially useful with mineral samples.
ü Dark-
field microscope: Bright
object, Dark background
The dark-field microscope produces detailed images of
living, unstained cells, and organisms by simply changing the way in which they
are illuminated. A hollow cone of light is focused on the specimen in such a
way that unreflected and unrefracted rays do not enter the objective. The only light that has been reflected or refracted by the specimen forms an image. The
field surrounding a specimen appears black, while the object itself is brightly
illuminated. The dark-field microscope can reveal considerable internal
structure in larger eukaryotic microorganisms. It also used to identify certain
bacteria such as the thin and distinctively shaped Treponema pallidum, the causative agent of syphilis.
Working
·
Light
enters the microscope for illumination of the specimen. A specially sized disc,
the patch stop or dark-field stop, blocks some light from the light source,
leaving an outer ring of illumination. A wide phase annulus can also be
reasonably substituted at low magnification.
·
The condenser lens focuses the light on the specimen. The light enters the
specimen. Most are directly transmitted, while some is scattered from the
specimen.
·
The
scattered light enters the objective lens, while the directly transmitted light
simply misses the lens and is not collected due to a direct-illumination block.
·
Only
the scattered light goes on to produce the image, while the directly
transmitted light is omitted.
Dark-field microscopy- In dark-field microscopy, a dark-field stop (inset)
is placed underneath the condenser lens system. The condenser then produces a
hollow cone of light so that the only light entering the objective comes from
the specimen.
Advantages
·
A
dark field microscope is ideal for viewing objects that are unstained,
transparent and absorb little or no light.
·
These
specimens often have similar refractive indices as their surroundings, making
them hard to distinguish with other illumination techniques. But in darkfield, the quality of images obtained is impressive and having high contrast. Darkfield is used to study marine organisms such as algae and plankton, diatoms,
insects, fibers, hairs, yeast, and protozoa as well as some minerals and
crystals, thin polymers, and some ceramics.
·
Dark the field is also used in the research of live bacterium, as well as mounted cells and
tissues.
·
It
is more useful in examining external details, such as outlines, edges, grain
boundaries and surface defects than internal structure.
Disadvantages
·
First,
dark field images are prone to degradation, distortion, and inaccuracies.
·
The specimen is strongly illuminated, which can cause damage to the specimen.
·
The specimen that is not thin enough or its density differs across the slide, may
appear to have artifacts throughout the image.
·
The
preparation and quality of the slides can grossly affect the contrast and
accuracy of a dark field image.
·
It
need special care that the slide, stage, nose and light source are free from
small particles such as dust, like these, will appear as part of the image.
·
Similarly,
if need to use oil or water on the condenser and/or slide, it is almost
impossible to avoid all air bubbles. These liquid bubbles will cause images
degradation, flare and distortion and even decrease the contrast and details of
the specimen.
·
Dark the field needs an intense amount of light to work. This, coupled with the fact
that it relies exclusively on scattered light rays, can cause glare and
distortion.
·
It
is not a reliable tool to obtain accurate measurements of specimens.
·
Finally,
numerous problems can arise when adapting and using a dark field microscope.
The amount and intensity of light, the position, size, and placement of the
condenser and stop need to be correct to avoid any aberrations.
·
However,
when employing this technique as part of a research study, need to take into consideration
the limitations and knowledge of possible unwanted artifacts.
Great keep it up...
ReplyDeleteVery useful content ritika... Best of luck for upcoming blogs.. 👍😘
ReplyDelete👍
ReplyDelete👍
ReplyDeleteHelpfull...
ReplyDeleteVery informative notes of microscopy .......keep it up 😘😘😘😘😘😘
ReplyDeleteOsm
ReplyDeleteInformative
ReplyDelete